ANNEX PUBLISHERS

Journal of Gynecology Research

ISSN: 2454-3284

Open Access
Research Article
Max Screen >>

A New Model of Obstetrical Pessary – Prevention and Treatment of Cervical Insufficiency and Habitual Abortion

Received Date: September 01, 2016 Accepted Date: November 15, 2016 Published Date: November 18, 2016

Copyright: © 2016 Schneiderman M. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Related article at Pubmed, Google Scholar

Abstract

This article presents a new model of obstetrical pessary for prevention and treatment of cervical insufficiency and habitual abortion.

In 2012 Research Center for Obstetrics, Gynecology and Perinatology (of the Ministry of Health of Russia) received an assignment from the Ministry of Health of Russia to develop a new medical device for correction and treatment of conditions associated with habitual abortion. In accordance with this task, a group of the Center`s researchers and obstetricians developed a new model of obstetrical pessary. In 2013 the Ethics Commission on biological and medical investigations of the Center authorized clinical use of this device.

Keywords: Obstetrical pessary; Cervical insufficiency; Habitual abortion; Cervical length; Preterm birth

Introduction

For many years the following obstetrical pessaries have been used in Russia: “Yunona” by the Simurg, Belarus, and “Dr. Arabin”, manufactured in Germany [1,2].

Review of existing models of pessaries has identified a number of shortcomings in their design and adverse effects or complications during use. The most common of these are discomfort, pain, expulsion, vaginal irritation, bacterial vaginosis, actinomycosis and allergic reactions to latex inducing discontinuation. These adverse events are associated with inadequate form or material of a pessary. Excessive vaginal secretion and erosions are also the sequela of the inadequate quality of the material used, its density, susceptibility to microbial contamination and chemical degradation [3,4]. The main objective was therefore to design a new model of obstetrical pessary without the above-mentioned shortcomings.

Methods

A new model of obstetrical pessary designed by the authors and developed jointly by the Research Center for Obstetrics, Gynecology and Perinatology of the Ministry of Health of Russia and ZAO “MEDSIL” is made of high quality silicone of predefined flexibility and density providing optimal application of the pessary. Application of pessary is easy and painless.

A specific feature of the new model of obstetrical pessary by Dr. Shneiderman (Model 1/Figure 1) are four semi-circular symmetrically located notches on the outer surface of the ring (due to these the pessary adjoins the vaginal walls and is retained reliably in a predetermined position without displacement). Another useful feature is four symmetrical semicircular bulges on the inner surface of the ring, which compress the cervix from four sides, preventing its opening (Figure 1) [5].

An exclusive innovation of pessary (Model 2/Figure 2) are four vertical silicon “lobes” at the lower surface of pessaries preventing from a shift or turn to a wrong position.

Absolutely new and significant advantage of the new model is its antimicrobial coating (miramistin, chlorhexidine, silver), preventing development of bacterial vaginosis which may impede use of a pessary [6].

The mechanism of prevention of cervical dilatation is based on cervical compression provided by the internal bulges and partial transfer of intrauterine pressure to the anterior wall of the uterus.

The inner part of the pessary is made of silicone rubber with Shore hardness A-60 units, and the outer part - of silicone rubber with Shore hardness of A–10 units or of cellular rubber, containing 2-4 wt. of silver nanoparticles.

The outer layer of silicone rubber with hardness of A-10 units or of cellular rubber applies minimal pressure to adjacent tissues, allowing avoiding their irritation, pressure ulcers, erosion, and stretching. In the same time, the use for the inner rubber layer with a Shore hardness of A-60 units provides the necessary compression and limits axial and radial movement of the device [7].

A new modified model has several structural differences from existing devices and reliably prevents the opening of the cervix in pregnant women with cervical incompetence as well as stress urinary incontinence. Installation of pessary is possible in outpatient setting thus providing the economic feasibility of the treatment.

Installation of the pessary should be done by an obstetrician-gynecologist. The doctor holds the pessary (its 2 opposite outer cuts) with two fingers, compressing it in vertical position, and introduces it into vagina up to its vault. The pessary can be easily and painlessly removed from vagina (Figure 3).

Then with the index finger the doctor ensures that the cervix comes to the central opening of the pessary, and the installation is accomplished.

The pessary is available in two sizes: for parous and nulliparous women. The appropriate choice cancels use of diaphragm in case if a pessary does not fit (Figure 4).

This study included 108 pregnant women aged 18 - 43 years (average 20.4 ± 1.2 years). All patients underwent clinical and laboratory evaluation and ultrasonic monitoring of the cervical internal os and cervical canal, defining the degree of severity of cervical insufficiency. Cervical incompetence was classified according to the following categories:

1. Congenital isthmic-cervical insufficiency (due to uterine malformations, genital infantilism).
2. Functional cervical insufficiency (due to endocrine dysfunction).
3. Organic isthmic-cervical insufficiency (following traumatic birth, manipulations or surgery on the cervix).

The optimal terms for use of obstetrical pessary are 17-37 weeks of gestation.

Results

The study of the new model of obstetrical pessary by doctor Schneiderman demonstrated its high reliability and efficiency. The patients` personal data and pregnancy outcomes after the treatment of isthmic-cervical insufficiency using pessary are presented in Table A.

The subjects of the study were 108 women diagnosed with isthmic-cervical insufficiency, followed for three years. The criteria for initiation of the preventive treatment of patients at high risk of preterm birth were the following: cervical length ≤ 25 mm measured by ultrasonography and dilatation of cervical internal os more than 0,8 mm at the 15th – 16th week of gestation and every 2 weeks on wards.

After installation of the pessary 2 patients (1.16%) reported the sensation of “fullness” in vagina, which resolved in one-two days. 34 women had the history of miscarriage in the 2nd trimester, 34 women (31.4%) were nulliparous, 8 women (7.3%) had previous operations on the cervix. Cervical length in 32 women (29.6%) was less than 2.5 cm. The pessaries were installed within the period 17.3 - 31.2 weeks of gestation. 99 women (91.6%) delivered after the 34th week of gestation, 9 women (8.3%) delivered before the 34th week. The outcomes of all deliveries were successful. There was not a single case of development of pressure sores of the vaginal wall or bleeding associated with trauma to the vaginal mucosa. There were no uterine contractions before or after installation of the pessary, therefore no tocolysis was necessary. Removal of the pessary at the 37th – 38th weeks of gestation was painless and uneventful. The use of the new pessary allowed to maintain pregnancy in all 108 pregnant women to term [8,9]. The analysis of pregnancy duration after treatment with pessary is demonstrated in Table A.

The majority of patients using pessaries delivered at 39 - 40 weeks though the pessaries were removed at 37th – 38th weeks of gestation and gestation was prolonged by 8-12 days before labor. Urogenital infection associated with isthmic-cervical insufficiency was diagnosed in 10.4% of cases [10]. Bacterial vaginosis, candidiasis and opportunistic infections were mostly detected. Analysis of complications during pregnancy, delivery and postpartum period is shown in Table B.

Low incidence or absence of severe obstetrical complications such as intrauterine growth restriction and placental abruption. Chorioamnionitis was not detected.

Discussion

The article represents study results of the use of new models of obstetrical pessaries designed by Dr. Schneiderman in pregnant women with isthmic-cervical insufficiency and habitual abortion. Different aspects of the use of the new models of obstetrical pessaries have been investigated. The criteria for the treatment of patients at high risk of miscarriage or preterm birth were the following: cervical length < 25 mm and dilatation of internal cervical os more than 0.8 mm. The use of the new pessaries enabled to prolong pregnancy by 45 (18-62) days.

The advantages of the new models of the obstetrical pessary are as follows:

1. Bulges on the internal surface of the ring preventing cervical dilatation in case of cervical insufficiency during pregnancy.
2. Stable fixation of the pessary in the vagina due to the external slots and lobes (preventing displacement or expulsion).
3. Reduced risk of lacerations of the vaginal mucosa due to minimal pressure to the vaginal wall.
4. Facilitated outflow of the vaginal discharge.
5. High comfort and easy installation.
6. Allergy free pessary.
7. Easy and painless removal of the pessary.
8. Elimination of symptoms of stress urinary incontinence.

Investigation of numerous cases evidences that the use of the new models of obstetrical pessaries provides support of pregnancy in patients with isthmic-cervical insufficiency.

Lack of comparative analysis of different models of pessaries is a drawback of this investigation; however again, the aim of the present study was to investigate a possibility of use of the new models of pessaries in clinical practice. In future we are planning to carry out detailed comparative analysis using different models of currently used pessaries.

Conclusion

The major challenge of prevention of preterm birth and continuation of the pregnancy puts the task to manage isthmic-cervical insufficiency and recurrent miscarriage.

Application of the new model of obstetrical pessary considerably increases the chances of successful maintenance of pregnancy course to term in women with cervical insufficiency and habitual abortion and reduces negative health outcomes.

5 Schneiderman M (2014) Patent N142309 Obstetrical pessary Recorded into the state registry of utility models of the Russian Federation.
7 Gorshkov AВ, Schneiderman MG, Zhukova EE, Biber BL, Valetova NR, Orlova OE, Kosinova TE Patent for the utility model N138859 Urogynecological device
Authors:
Recorded into the state registry of utility models of the Russian Federation 28 February 2014
Materials from the XV Russian scientific forum “Mother and Child”, Moscow, 23 September 2014.

Journal of Gynecology Research

Tables at a glance
Table A
Table B
Figures at a glance
Figure 1
Figure 2
Figure 3
Figure 4

Figure 1: Obstetrical pessary by Dr. Shneiderman (Model 1: four symmetrical semicircular slots on the external surface of the ring, four bulges on the internal surface of the ring.)

Figure 2: Obstetrical pessary by Dr. Shneiderman (Model 2)

Figure 3: Installation of the pessary

Figure 4: Pessaries for parous (left) and nulliparous (right) women

Age (average)
20.4 + 1.2 (18-43)
One miscarriage in the 2nd trimester of pregnancy
22 (20.3%)
Two miscarriages in the 2nd  trimester of pregnancy
10 (9.25%)
Three miscarriages in the 2nd  trimester of pregnancy
2 (1.85%)
Nulliparous
34 (31.4%)
Preterm birth
46 (42.6%)
Cervical surgery
8 (7.3%)
Average cervix (cm)
2.0 (1.6-2.5)

Table А: Medical history data of patients

Terms of pregnancy at pessary installation (average, weeks)
24.1 (17.1-31.2)
Discomfort after installation of the pessary (1-2 days)
2 (1.16%)
Interval from removal of the pessary to delivery (days)
10.0 (8-12)
Term of delivery (average, weeks)
36.7 (34-40.2)
Delivery after the 34th week of gestation
99 (91.6%)
Delivery before the 34th week of gestation
9 (8.3%)
Duration of pregnancy prolongation (days)
45 (18.0-62.0)
Cervical length <2.5cm
32 (29.6%)
Cervical length >2.5cm
76 (70.3%)
Dilatation of internal cervical os (>0.8mm)
64 (59.2%)

Table B: Outcomes